    Cardiac resynchronization therapy restores optimal atrioventricular mechanical timing in heart failure patients with ventricular conduction delay

    Get PDF
    AbstractObjectivesWe characterized the relationship between systolic ventricular function and left ventricular (LV) end-diastolic pressure (LVEDP) in patients with heart failure (HF) and baseline asynchrony during ventricular stimulation.BackgroundThe role of preload in the systolic performance improvement that can be obtained in HF patients with LV stimulation is uncertain.MethodsWe measured the maximum rate of increase of LV pressure, LVEDP, aortic pulse pressure (PP) and the atrioventricular mechanical latency (AVL) between left atrial systole and LV pressure onset in 39 patients with HF. Two subgroups were identified: “responder” if PP improved, or “nonresponder.”ResultsMaximum hemodynamic improvement occurred at an atrioventricular (AV) delay that did not decrease LVEDP. Left ventricular and biventricular (BV) stimulation increased systolic hemodynamics significantly, despite no significant increase in LVEDP. All parameters decreased when the LVEDP was decreased by shorter AV delay. Left ventricular and BV stimulation provided better hemodynamics than right ventricular (RV) stimulation. For the nonresponder subgroup, systolic hemodynamics only worsened during AV delay shortening. For the responder subgroup, optimum PP was achieved when AVL was near zero.ConclusionsRestoration of optimal left atrial-ventricular mechanical timing partly contributes to the hemodynamic improvements observed in this patient subgroup. However, preload alone cannot explain the differences seen between RV and BV stimulation and the contradictory PP decreases even at maximal preload in the nonresponder subgroup. These results may be explained by a site-dependent mechanism such as the degree of ventricular synchrony. Caution should be taken in these patients when optimizing AV delays using echocardiography techniques that focus on LV inflow

    Chronic Low-Level Vagus Nerve Stimulation Improves Long-Term Survival in Salt-Sensitive Hypertensive Rats

    Get PDF
    Chronic hypertension (HTN) affects more than 1 billion people worldwide, and is associated with an increased risk of cardiovascular disease. Despite decades of promising research, effective treatment of HTN remains challenging. This work investigates vagus nerve stimulation (VNS) as a novel, device-based therapy for HTN treatment, and specifically evaluates its effects on long-term survival and HTN-associated adverse effects. HTN was induced in Dahl salt-sensitive rats using a high-salt diet, and the rats were randomly divided into two groups: VNS (n = 9) and Sham (n = 8), which were implanted with functional or non-functional VNS stimulators, respectively. Acute and chronic effects of VNS therapy were evaluated through continuous monitoring of blood pressure (BP) and ECG via telemetry devices. Autonomic tone was quantified using heart rate (HR), HR variability (HRV) and baroreflex sensitivity (BRS) analysis. Structural cardiac changes were quantified through gross morphology and histology studies. VNS significantly improved the long-term survival of hypertensive rats, increasing median event-free survival by 78% in comparison to Sham rats. Acutely, VNS improved autonomic balance by significantly increasing HRV during stimulation, which may lead to beneficial chronic effects of VNS therapy. Chronic VNS therapy slowed the progression of HTN through an attenuation of SBP and by preserving HRV. Finally, VNS significantly altered cardiac structure, increasing heart weight, but did not alter the amount of fibrosis in the hypertensive hearts. These results suggest that VNS has the potential to improve outcomes in subjects with severe HTN

    Central-Peripheral Neural Network Interactions Evoked by Vagus Nerve Stimulation: Functional Consequences on Control of Cardiac Function

    No full text
    Using vagus nerve stimulation (VNS), we sought to determine the contribution of vagal afferents to efferent control of cardiac function. In anesthetized dogs, the right and left cervical vagosympathetic trunks were stimulated in the intact state, following ipsilateral or contralateral vagus nerve transection (VNTx), and then following bilateral VNTx. Stimulations were performed at currents from 0.25 to 4.0 mA, frequencies from 2 to 30 Hz, and a 500-μs pulse width. Right or left VNS evoked significantly greater current-and frequency-dependent suppression of chronotropic, inotropic, and lusitropic function subsequent to sequential VNTx. Bradycardia threshold was defined as the current first required for a 5% decrease in heart rate. The threshold for the right vs. left vagus-induced bradycardia in the intact state (2.91 ± 0.18 and 3.47 ± 0.20 mA, respectively) decreased significantly with right VNTx (1.69 ± 0.17 mA for right and 3.04 ± 0.27 mA for left) and decreased further following bilateral VNTx (1.29 ± 0.16 mA for right and 1.74 ± 0.19 mA for left). Similar effects were observed following left VNTx. The thresholds for afferent-mediated effects on cardiac parameters were 0.62 ± 0.04 and 0.65 ± 0.06 mA with right and left VNS, respectively, and were reflected primarily as augmentation. Afferent-mediated tachycardias were maintained following β-blockade but were eliminated by VNTx. The increased effectiveness and decrease in bradycardia threshold with sequential VNTx suggest that 1) vagal afferents inhibit centrally mediated parasympathetic efferent outflow and 2) the ipsilateral and contralateral vagi exert a substantial buffering capacity. The intact threshold reflects the interaction between multiple levels of the cardiac neural hierarchy

    Aberrant Fecal Flora Observed in Guinea Pigs With Pressure Overload Is Mitigated in Animals Receiving Vagus Nerve Stimulation Therapy

    No full text
    Altered gut microbial diversity has been associated with several chronic disease states, including heart failure. Stimulation of the vagus nerve, which innervates the heart and abdominal organs, is proving to be an effective therapeutic in heart failure. We hypothesized that cervical vagus nerve stimulation (VNS) could alter fecal flora and prevent aberrations observed in fecal samples from heart failure animals. To determine whether microbial abundances were altered by pressure overload (PO), leading to heart failure and VNS therapy, a VNS pulse generator was implanted with a stimulus lead on either the left or right vagus nerve before creation of PO by aortic constriction. Animals received intermittent, open-loop stimulation or sham treatment, and their heart function was monitored by echocardiography. Left ventricular end-systolic and diastolic volumes, as well as cardiac output, were impaired in PO animals compared with baseline. VNS mitigated these effects. Metagenetic analysis was then performed using 16S rRNA sequencing to identify bacterial genera present in fecal samples. The abundance of 10 genera was significantly altered by PO, 8 of which were mitigated in animals receiving either left- or right-sided VNS. Metatranscriptomics analyses indicate that the abundance of genera that express genes associated with ATP-binding cassette transport and amino sugar/nitrogen metabolism was significantly changed following PO. These gut flora changes were not observed in PO animals subjected to VNS. These data suggest that VNS prevents aberrant gut flora following PO, which could contribute to its beneficial effects in heart failure patients

    Defibrillation efficacy of different electrode placements in a human thorax model

    No full text
    The objective of this study was to measure the defibrillation threshold (DFT) associated with different electrode placements using a three- dimensional anatomically realistic finite element model of the human thorax. Coil electrodes (Endotak DSP, model 125, Guidant/CPI) were placed in the RV apex along the lateral wall (RV), withdrawn 10 mm away from the RV apex along the lateral wall (RVprox), in the RV apex along the anterior septum (RVseptal), and in the SVC. An active pulse generator (can) was placed in the subcutaneous prepectoral space. Five electrode configurations were studied: RV→SVC, Rv(prox)→SVC, RV(SEPTAL)→SVC, RV→Can, and RV→SVC+Can. DFTs are defined as the energy required to produce a potential gradient of at least 5 V/cm in 95% of the ventricular myocardium. DFTs for RV→SVC, RV(prox) →SVC, RV(septal)→SVC, RV→Can, and RV→SVC+Can were 10, 16, 7, 9, and 6J, respectively. The DFTs measured at each configuration fell within one standard deviation of the mean DFTs reported in clinical studies using the Endotak leads. The relative changes in DFT among electrode configurations also compared favorably. This computer model allows measurements of DFT or other defibrillation parameters with several different electrode configurations saving time and cost of clinical studies